57
Bruker Optik GmbH HYPERION User Manual
Operation 4
Important note: After each reboot of the PC, you have to switch off the FPA detec-
tor and then to switch on again. Before switching on the FPA detector, make sure
that the FPA detector is cooled down to its operating temperature. For information
about how to cool down the FPA detector, see section 4.4.
3. Cooling down the FPA detector with liquid nitrogen to its operating tempera-
ture
See section 4.4
4. Switching on the PC and the monitor
For information about this topic, refer to the corresponding user manuals.
5. Switching on the FPA detector
The ON/OFF switch
is at the bottom
side of the FPA detector compartment
Switching off the analysis system
Note: When switching off the analysis system, there is no strict sequence of
actions you have to observe.