Operation 4
In measuring mode
Proceed as follows:
1
Bring the ATR crystal in the measuring mode position (i.e. lower position).
☞ See section 4.12.4.
2
Set a contact pressure level which is adequate to the degree of hardness of the
sample you intend to analysis.
☞ See section 4.12.5.
3
Focussing for the background mea-
surement
Put the plastic ring
① on the stage.
Position the ring concentrically under-
neath the ATR objective. Make sure
that the ring does not yet contact the
ATR objective.
Note: The plastic ring is included in the
delivery scope of the ATR objective.
Focussing for the sample measure-
ment
Put the sample on the stage under-
neath the ATR objective.
4
Focus by moving the stage slowly
upwards until the ATR crystal is
pressed in the focus position by plastic
ring or the sample. In case of a man-
ual stage or a motorized x/y-stage, use
the fine focus knobs (C/C’ in fig. 3.7).
In case of a motorized x/y/z stage, use
the joystick.
Note: The focus position is indicated
acoustically by a beep and optically by
the In-focus LED
②. When the ATR
crystal has reached the focus position,
the In-focus LED turns from red to
green for a short moment. Then it
goes off.
Red: ATR crystal is not in focus posi-
tion.
Green: ATR crystal is in focus position.
Note: In case the ATR crystal is in the
viewing mode position, the In-focus
LED does not light at all.