Bruker HYPERION - Measuring Sample with ATR Objective (Mapping)

Bruker HYPERION
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
93
Bruker Optik GmbH HYPERION User Manual
Operation 4
13. Start background measurement. OPUS software
See OPUS/VIDEO Manual.
14. Move the stage downwards as far
as it will go. Exchange the plastic ring
for the sample.
Manual stage and motorized x/y-
stage: coarse and fine focus knobs at
the microscope (B/B’ and C/C’ in fig. 3.7)
Motorized x/y/z-stage: Joystick and cor-
responding stage control functions in
OPUS
See OPUS/VIDEO Manual.
15. Bring the ATR crystal in the view-
ing mode position (i.e. upper position).
ATR objective
See section 4.12.4.
16. Activate the viewing mode at the
microscope.
either the VIS button on the operating
panel (B in fig. 3.4) or in the OPUS video
wizard
See OPUS/VIDEO Manual.
17. Focus on the sample by moving
the stage slowly upwards.
Manual stage and motorized x/y-
stage: coarse and fine focus knobs at
the microscope (B/B’ and C/C’ in fig. 3.7)
Motorized x/y/z-stage: Joystick and cor-
responding stage control functions in
OPUS
See OPUS/VIDEO Manual.
18. Find a sample area which you
intend to analyze spectroscopically by
moving the stage in x- and/or y-direc-
tion.
Manual stage: Rotary knobs (A and B in
fig. 3.6)
Motorized stage: Joystick and corre-
sponding stage control functions in
OPUS
See OPUS/VIDEO Manual.