3
MicroDock II
Safety Information - Read First
a Cautions
MicroDock ll base station
• If the base station is damaged or parts are missing, contact BW Technologies by Honeywell
immediately.
• The base station must be used only in an area that is free of hazardous gas. Do not use the base station in a hazardous
area. Failure to adhere to this caution can lead to fire and/or explosion.
• This equipment uses potentially harmful gas for calibrations. The base station must be attached to a venting system or
be used in a well-ventilated area.
• Do not calibrate the detector during or immediately after charging is complete.
• Perform calibrations and bump tests only in a safe area that is free of hazardous gas, in an atmosphere of 20.9% oxy-
gen.
• The maximum recommended exhaust line length is 50 ft. (15.24 m).
• Ensure that the inlet filter is clean.
• Ensure that all gas cylinders contain enough gas.
• Ensure the exhaust line is not connected to a negative pressure system.
• A demand flow regulator must be used with all gas cylinder connections. Input pressure not to exceed 10 psi.
• Do not expose the base station to electrical shock or severe continuous mechanical shock.
• The base station warranty will be void if the unit is disassembled, adjusted, or serviced by non-BW Technologies by
Honeywell personnel.
• Do not immerse the base station in liquids.
GasAlertMicro, GasAlertMicro 5/PID/IR, GasAlertMicroClip, GasAlertMax XT, and GasAlertQuattro Chargers
• If the charger is damaged or parts are missing, contact BW Technologies by Honeywell
immediately.
• Do not change or charge the batteries in a hazardous location. Do not use the charger in a hazardous location. Failure
to adhere to these cautions can lead to fire and/or explosion.
• Read and adhere to all instructions and cautions that are provided with the charger. Failure to do so can result in fire,
electric shock, personal injury, and/or property damage.