MicroDock II
Order of Gases Applied for Bump Tests and Calibrations
Order of Gases Applied for Bump Tests and
Calibrations
a Warning
It is extremely important that gases are applied in the order that is
specified in this manual.
Failure to adhere to the rules in Table 19.
and Table 20. of this
manual will result in incorrect sensor readings and possible
personal injury and/or property damage.
Note
Manual application of gas types is available for the
GasAlertMicro 5/PID/IR detectors only.
To manually apply gases for bump tests and calibrations, complete the
following:
1. Review Table 19.
and Table 20.
2. Determine which gases will be applied.
3. Using Table 19.
, locate the first gas type (sensor) to be bump
tested or calibrated on the top row.
4. Within the far left column, locate the next gas type (sensor) to be
bumped or calibrated.
The square where the column and row intersect will indicate
which gas should be applied first and a rule number (if applica-
ble).
5. Using Table 20.
, locate the corresponding rule number. Ensure
that the desired order to apply gases corresponds to the gas
application rules.
6. Continue to reference Table 19.
and Table 20. to ensure the
gases are applied in the correct order.
7. When the correct order is determined, proceed to the Bump Test
or Calibration
section.