EasyManua.ls Logo

BW Technologies MicroDock II - Gas Conflicts

BW Technologies MicroDock II
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MicroDock II
User Manual
54
20. Successful Reconfiguring: If reconfiguration is successful, the
following screen displays.
Unsuccessful Reconfiguration: If reconfiguration is not suc-
cessful, the following screen displays.
Press
C ok. The LCD then returns to normal operation.
Resolving IR Errors: Refer to the following solutions. If the following
solutions do not resolve the IR error, refer to Troubleshooting
.
Check the lighting conditions. Infrared or intense light (sunlight or
halogen) can cause IR disruptions.
Remove and reposition the detector in the docking module.
Communication between the detector and base station may have
been temporarily disrupted. Complete the procedures again.
Gas Conflicts
When performing a bump test or calibration, gas must be applied in a
specific order to prevent gas conflicts that can damage the detector’s
sensor(s).
Note
Ensure that the gas inlets are configured correctly. Refer to
Inlet Select
.
The gas conflicts screen automatically displays if a gas conflict occurs
when a bump check or calibration is initiated.
There are two types of gas conflicts:
Conflicts: More than one inlet is configured for a specific gas
type.
Example: Inlet 2 is configured for SO
2
and inlet 3 is configured
for the 3-gas SO
2
mix.
Not Found: The base station is unable to locate the required gas
type for a specific sensor on the detector.
The base station displays additional information regarding the
number of gas conflicts,
number of gases not found,
docking module (e.g., Bay 1),
detector gas type(s),
inlet, and
inlet gas type(s).

Table of Contents

Other manuals for BW Technologies MicroDock II