Bump Test
Bump Test Results
After the bump test is complete, the base station displays the results of
the test.
Bump Pass
indicates that the detector passed the bump tests. The detector is
now ready for use. Press
C OK to return to normal operation.
Bump Fail
If the detector failed a bump test, displays beside the applicable gas
Example: H2S .
If the detector fails an audible or visual test, displays beside the appli-
cable test (AUD or VIS). For information regarding solutions, refer to
Troubleshooting
.
Note
If a gas type fails, calibrate the detector prior to use.
Confirm that the gas concentration exceeds the low alarm set-
point of the detector being tested (applicable to GasAlertClip
Extreme, GasAlert Extreme, and GasAlertMicro only).
Confirm the inlets are setup correctly.
Pass
Fail
H2S
Gas type not detected (H
2
S is used as an
example, this applies to all gases)
Sensor is disabled