71
MicroDock II
Calibration
Incorrect Inlet Setup
If an inlet is not setup correctly during a calibration, the base station dis-
plays Fail and a hyphen ( ) beside the gas type.
Example: H
2
S, CO, and LEL gases were not detected.
If a test gas fails, verify that the
• gas cylinder is not empty,
• connections are attached correctly, and
• inlets are setup correctly.
If the detector fails an audible or visual test, displays beside the appli-
cable test (
AUD or VIS). For information regarding solutions, refer to
Troubleshooting
.
Sensor Disabled
a Warning
To prevent personal injury and/or property damage, enable or
replace the sensor immediately. Refer to the corresponding
detector user manual for sensor replacement procedures.
If a sensor is disabled, results from the sensor do not display on the base
station LCD. displays where the gas type typically displays on the
LCD.
Note
If a sensor is disabled, it will not be recorded in the results log.