EasyManuals Logo

Cardinal 201 User Manual

Cardinal 201
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
Model 201 Installation & Technical
8400-M117-O1 Rev D
Page 36
CAL3= Third Calibration Weight
1. The display will show CAL3=. This is the third of three
calibration weights. This weight could be ZERO (NO LOAD)
or the TEST WEIGHTS (TEST LOAD).
2. Press the F3/ key to view the current setting.
3. If the third calibration weight is to be ZERO (NO LOAD),
press the F3/ key.
4. If the third calibration weight is to be the TEST WEIGHTS
(TEST LOAD), use the F1/ and F2/ keys to input the
value of the test weights.
5. Place the weights on the scale platform, then press the F3/
key.
6. Starting at the left and proceeding right, a series of dashes
will appear on the display.
7. Next, starting at the left and proceeding right, the dashes will
disappear, after which the display will show: F SPAN.

Table of Contents

Other manuals for Cardinal 201

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cardinal 201 and is the answer not in the manual?

Cardinal 201 Specifications

General IconGeneral
BrandCardinal
Model201
CategoryTransmitter
LanguageEnglish