Model 201 Installation & Technical
8400-M117-O1 Rev D
Page 36
CAL3= Third Calibration Weight
1. The display will show CAL3=. This is the third of three
calibration weights. This weight could be ZERO (NO LOAD)
or the TEST WEIGHTS (TEST LOAD).
2. Press the F3/ key to view the current setting.
3. If the third calibration weight is to be ZERO (NO LOAD),
press the F3/ key.
4. If the third calibration weight is to be the TEST WEIGHTS
(TEST LOAD), use the F1/ and F2/ keys to input the
value of the test weights.
5. Place the weights on the scale platform, then press the F3/
key.
6. Starting at the left and proceeding right, a series of dashes
will appear on the display.
7. Next, starting at the left and proceeding right, the dashes will
disappear, after which the display will show: F SPAN.