Model 201 Installation & Technical
8400-M117-O1 Rev D
4.7 Calibration
CAL
With CAL displayed, press the F3/ key. The display will change to CAL=.
Proceed to the CAL= (Perform Calibration) parameter.
CAL= (Perform Calibration)
With CAL= displayed, press the F3/ key. The display will change to
no. If scale has been previously calibrated and you wish to skip
calibration and proceed to F SPAn, press the F3/ key again.
Otherwise, press the F2/ key to toggle to yes and then press the F3/
key. The display will change to CAL1=. Proceed to the CAL1=
parameter.
During calibration it will be necessary to enter values using the 201
keypad.
Pressing the F3/ key will cause the data entered or displayed to be
retained and the 201 to advance to the next prompt.
The functions of numeric keys are replaced by using the F1/ and the
F2/ keys.
The cursor location is identified by the blinking character and can be
advanced to the left to the next position by pressing the F1/ key.
Pressing the F2/ key will change the blinking character to the next
value.