Model 201 Installation & Technical
8400-M117-O1 Rev D
Page 89
5.13 Enable Flow Rate
FLo
With FLo displayed, press the F3/ key. The display will change to Flo=.
Proceed to the Flo= (Enable Flow Rate Measuring) parameter.
Flo= (Enable Flow Rate Measuring)
Press the F3/ key to show the current setting. If the setting displayed is
acceptable, press the F3/ key again to save it. Otherwise, use the
F2/ key to toggle to a new setting and then press the F3/ key to save
it. Allowable settings are:
no
no (Flow Rate Measuring is Disabled)
YES
YES (Flow Rate Measuring is Enabled)
NOTE: If Flo= (Flow Rate Measuring) is Enabled (YES), the
following prompt, Unit will be active.
Unit= (Units for weight change)
Press the F3/ key to show the current setting. If the setting displayed is
acceptable, press the F3/ key again to save it. Otherwise, use the
F2/ key to toggle to a new setting and then press the F3/ key to save
it. Allowable settings are:
10tH S
Weight Change per 1/10th Second
seC
Weight Change per Second

Weight Change per Minute
Hour
Weight Change per Hour