5. WEIGHT TRANSMITTER SETUP (Non-Metrological Parameters)
The Model 201 allows several non-metrological parameters to be
reviewed and changed without breaking the calibration seal.
5.1 Accessing Non-Metrological Parameters
1. With the weight transmitter ON, press the F1/ key and the F2/ key
simultaneously.
2. The display will change to show dAte (the prompt to set the date
and time).
3. Press the F3/ key to show the current setting. If the setting
displayed is acceptable, press the F3/ key again to save it.
Otherwise, use the F2/ key to toggle to a new setting and then
press the F3/ key to save it and advance to the next prompt.
4. To exit press the F3/ key to step through the remaining prompts
OR at anytime, cycle the power.
5. The weight transmitter will return to the normal weight display.
Event Counter (Read Only Values)
Accumulators and Consecutive Number
Preferences (Date order, 12 or 24 hour clock , Function
key assignments and default background Color
Analog Input/Output Setup
Digital Input/Output Setup
Optional Memory Card Configuration
Check weighing (over/under) Setup
Print Tab Settings Configuration
Display High Resolution Weight