Model 201 Installation & Technical
8400-M117-O1 Rev D
Page 27
4.5 Scale Configuration
SCALE
With SCALE displayed, press the F3/ key. The display will change to Unit1=.
Proceed to the Unit1= (Weighing Unit 1) parameter.
Unit1= (Weighing Unit 1)
With Unit1= displayed, press the F3/ key to show the current setting.
If the setting displayed is acceptable, press the F3/ key again to save it.
Otherwise, use the F2/ key to toggle to a new setting and then press the
F3/ key to save it. Allowable settings are:
1 lb = pounds
2 kg = kilograms
3 oz = ounces
4 g = grams
int= (Interval Setting)
Press the F3/ key to show the current setting.
If LFt=YES, use the F2/ key to toggle to a new setting and then press
the F3/ key to save it. Allowable settings are: 1, 2, 5, 10, 20 or 50.
If LFt=no, use the F2/ and F1/ keys to enter a new setting and then
press the F3/ key to save it. Allowable settings are: 1 through 99.
When the setting displayed is acceptable, press the F3/ key again to
save it.
dPP= (Decimal Point Setting)
Press the F3/ key to show the current setting. If the setting displayed is
acceptable, press the F3/ key again to save it. Otherwise, use the
F2/ key to toggle to a new setting and then press the F3/ key to save
it. Allowable settings are:
0 = X X X X X X
1= X X X X X. X
2 = X X X X. X X
3 = X X X. X X X
NOTE: The Decimal Point Setting prompt will not be shown if the Interval
Setting is 10 or greater.