CERUS INTERCEPT INT100 - • The correct dose of light is being administered to the blood product, so pathogen inactivation is being appropriately achieved. The treatment time automatically increases as the bulbs gradually become dimmer, so that the correct dose is administered.; • If the treatment time exceeds the appropriate range as set by the authorized service representative, a message will be displayed to change the bulbs.; • If the self-tests are not successful, an error message will appear on the screen with additional instructions. However, if these additional instructions are not helpful, turn the illuminator off, wait until the screen darkens, then turn it back on....; • If your operator badge will not scan into the illuminator, manually enter your operator username into the illuminator using the keypad.

CERUS INTERCEPT INT100
182 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 – Troubleshooting
Section 5.2 Summary of Questions and Answers
5 - 5
?
Q & A
What should you do if the treatment times in your illuminator are
consistently at the upper end of the range in Section 3.5?
The correct dose of light is being administered to the blood
product, so pathogen inactivation is being appropriately
achieved. The treatment time automatically increases as the
bulbs gradually become dimmer, so that the correct dose is
administered.
If the treatment time exceeds the appropriate range as set by
the authorized service representative, a message will be
displayed to change the bulbs.
Chapter 4: HOW TO USE THE ILLUMINATOR
? Q & A
What will happen if the self-tests are not successful
If the self-tests are not successful, an error message will appear on
the screen with additional instructions. However, if these
additional instructions are not helpful, turn the illuminator off, wait
until the screen darkens, then turn it back on. If this does not
resolve the issue, contact your authorized service representative
for assistance.
?
Q & A
What should you do if your operator badge will not scan into the
illuminator?
If your operator badge will not scan into the illuminator, manually
enter your operator username into the illuminator using the
keypad.
?
Q & A
How will you know if only one of the drawer chambers is working and can
be used?
When only one chamber is working and can be used, the
illuminator will display which chamber is available in the last
option of the menu on the ‘Number of containers’ screen. There
will be a blue ‘X’ on the screen across the chamber that is not
working and cannot be used. This is shown in the examples
below. Therefore, if chamber 1 is not working, you can use
chamber 2 to perform the illumination process. As you will not be
able to select two containers, the arrow buttons will not work at
this stage.
MAN-EN 00601, v3.0