CERUS INTERCEPT INT100 - • The correct dose of light is being administered to the blood product, so pathogen inactivation is being appropriately achieved. The treatment time automatically increases as the bulbs gradually become dimmer, so that the correct dose is administered.; • If the treatment time exceeds the appropriate range as set by the authorized service representative, a message will be displayed to change the bulbs.

CERUS INTERCEPT INT100
182 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 – Illuminator Description
Section 3.11 – Questions and Answers
3-55
?
Q & A
What should you do if the treatment times in your illuminator are
consistently at the upper end of the range in Section 3.5?
The correct dose of light is being administered to the blood
product, so pathogen inactivation is being appropriately
achieved. The treatment time automatically increases as the
bulbs gradually become dimmer, so that the correct dose is
administered.
If the treatment time exceeds the appropriate range as set
by the authorized service representative, a message will be
displayed to change the bulbs.
MAN-EN 00601, v3.0