Chapter 1 ā What the Illuminator Does
Section 1.7 ā Cautions
1-12
Chapter 6: MAINTENANCE, TRANSPORTATION AND STORAGE,
WARRANTY AND SERVICE
CAUTION
Do not use the illuminator if there is condensation on it.
Humidity higher than 80% may shorten the life of instrument
components.
Section 6.2 - Selecting a Place for the Illuminator
CAUTION
Use only approved solutions to clean and disinfect the tray;
otherwise, damage may occur.
Section 6.3 - Cleaning the Illuminator
CAUTION
If any part of the illuminator tray is damaged (cracked, scratched
or cloudy) contact your authorized service representative for a
replacement. Do not use the illuminator if the tray is damaged.
Section 6.3 - Cleaning the Illuminator
Chapter 7: SPECIFICATIONS
CAUTION
The illuminators need special precautions regarding
electromagnetic compliance (EMC) and need to be installed and
put into service according to the EMC information provided in
the above tables.
Section 7.2 - Illuminator Compliance with Standards
CAUTION
Portable and mobile radio frequency (RF) communications
equipment can affect the illuminators by interfering with the
electronics, causing the illumination to stop.
Section 7.2 - Illuminator Compliance with Standards
CAUTION
If the fuse blows after bulb replacement, the illuminator should
not be used. Contact your authorized service representative for
assistance.
Section 7.3 - Illuminator Requirements
CAUTION
If using Codabar (excluding Monarch 11 and UKBTS), the
illuminator will not evaluate the integrity of the scanned data.
Verify the data scanned matches the information displayed on
the illuminator screen.
Section 7.4 - Barcode Compatibility