EasyManua.ls Logo

Chroma 62000P Series - Page 78

Chroma 62000P Series
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Operation
Enter the voltage measured by DVM (4 digits after decimal point) and press “
ENTER
” to
confirm.
(8) Press
” again to do the high voltage range calibration for its second point, the
instrument will set the output voltage to 70.00V and the cursor is stopped at position [2]
as Figure 3-79 shows. Enter the voltage measured by DVM (4 digits after decimal point)
and press “
ENTER
” to confirm.
]3[
]4[
[VOLTAGE CALIB R A T I O N ]
R E M O V E E X T E R N A L L O A D A N D P R E S S [ E N T E R ]
S E T T I N G O U T P U T V O L T A G E 2 5 . 0 V
A C T U A L O U T P U T V O L T A G E = 2 4. 9 9 2 V
S E T T I N G O U T P U T V O L T A G E 7 0 . 0 V
A C T U A L O U T P U T V O L T A G E = 6 9 . 9 8 1 V
Figure 3-78
(9) Now the voltage calibration is done. To save the calibration data, press “
SAV E
” will
prompt a confirmation page as Figure 3-79 shows, press
ENTER
to save it. Press
EXIT
” to return to the Calibration page if not.
S
A
V
E=YES
Figure 3-79
(10) Press
EXIT
” to return to MAIN PAGE.
NOTICE
1. The calibration point may be different for other models (non 62012P-80-60), please
operate it following the instructions displayed.
3-49

Table of Contents

Related product manuals