EasyManuals Logo

Chroma 62000P Series User Manual

Chroma 62000P Series
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
Manual Operation
3.3.7.5.2 SETUP
123456
+12VA APIGND V SET I SET V
MEAS.
I
MEAS.
+
+
DC Power
Supply
DVM1 DVM2
+
Figure 3-100
3.3.7.5.3 Calibration Procedure (Example: Model 62012P-80-60)
(1) In CALIBRATION page, press “
4
” or turn “Rotary” (
) knob to set CHOICE = 4.
(2) Press
ENTER
” to confirm entering into APG Voltage Calibration options as Figure 3-101
shows.
[APG CU
R
RENT CAL I B
R
A
T I O N ]
C H E C
K
A P G C O N N E C T I O N A N D P
R
ESS [ E N T E
R
]
( S E T ) I N P U T V O L T A G E F O R S E T T I N G = 0 . 5
V
A
C T U
A
L A P G I N P U T V O L T A G E = 0 . 6 6 V
( S E T ) I N P U T V O L T A G E F O R S E T T I N G = 8 . 0
V
A
C T U
A
L A P G I N P U T V O L T A G E = 8 . 6 8 V
Figure 3-101
NOTICE
When entering into the CALILBRATION page, be sure to check the interface connection on
the rear panel is correct and then press “
ENTER
” to start calibration.
(3) When in the APG Current Calibration pages and the connection is correct, press “
ENTER
to confirm.
(4) It will ask users to input about 0.5V voltage signal (Pin 4), and press “
ENTER
” the cursor
will stop at position [1] as Figure 3-102 shows. Adjust the Power Supply to 0.5V±0.2V
and use DVM1 to measure the reading of Power Supply. Input the voltage (2 digits after
decimal point) read by DVM 1 to position [1] and press “
ENTER
” to confirm.
3-61

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 62000P Series and is the answer not in the manual?

Chroma 62000P Series Specifications

General IconGeneral
BrandChroma
Model62000P Series
CategoryPower Supply
LanguageEnglish

Related product manuals