Contact your local Cosmo representative for scheduling Annual Calibration Service.
The following items will be inspected and calibrated.
1) Check the oil mist separator and the filter.
2) Leak check of the leak tester
3) Check the DPS offset.
4) Check the DPS span
5) Check the PS offset.
6) Check the PS span.
4
4
F
F
e
e
a
a
t
t
u
u
r
r
e
e
s
s
f
f
o
o
r
r
M
M
a
a
i
i
n
n
t
t
e
e
n
n
a
a
n
n
c
c
e
e
4.1 K(Ve) Check
Compares K(Ve) measured with a reference master tested part to the K(Ve) stored in memory.
This can be used for daily sensitivity checks of Type J models.
K(Ve) Check behavior will be same as the K(Ve) Automatic Setup.
K(Ve) Check Limit
Setting a tolerance in percentage (±) to the K(Ve) currently stored in memory.
> Main Menu > K(Ve) > K(Ve) Settings > Basic > K(Ve) Check Limit
Manual Operation
1) > Main Menu > K(Ve) > K(Ve) Check
2) Press to start K(Ve) Check.
Remote Operation
Transmit K(Ve) Check signal and Start Signals through control I/O port.
Results for K(Ve) Check
Lower than LL Within the Limits Larger than UL
DET LL Pass DET UL

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cosmo LS-R700 and is the answer not in the manual?

Cosmo LS-R700 Specifications

General IconGeneral
BrandCosmo
ModelLS-R700
CategoryTest Equipment
LanguageEnglish

Related product manuals