6.1 Mastering Settings
Basic Settings
Disable/Enable Mastering Compensation feature
MB1 Timer BAL1 timer for Mastering: 0 to 999.9 [s]
Recommended timer: 1.0 [s]
MB2 Timer BAL2 timer for Mastering: 0 to 999.9 [s]
Recommended timer: 2.0 [s]
0 to ±9999.9 (programmed unit) Mastering Upper/Lower limits.
Normally 120 to 150% of DET UL/LL
Mastering Comp value can be entered automatically or automatically
Idle time in production line. Exceeding the time transmits Mastering
Request Signal. Setting this to 0 disable the feature.
Repeatedly transmits Mastering Request Signal at the beginning of the
programmed day of week for the programmed number of times with the
programmed interval.
Upper leak limit for transmitting Mastering Request Signal.
Over Comp (-) Limit
Lower leak limit for transmitting Mastering Request Signal.
Number of triggers for transmitting Mastering Request Signal.
Mastering Settings
Mastering Display
Fixed Comp Settings
Request Signal
MB1 Timer
MB2 Timer
M. Comp Lower Limit
Mastering Comp Value
Interval
Iteration
Consecutive Triggers
Refer to
“