9 Troubleshooting 115
3.9 ERROR 15 Air Operated Valve Error 5
At the end of Stabilization (BAL2) Only for High press and External Press.
Pressure switch monitoring the pilot pressure for the Balance (BAL) valve is not activated.
Pilot pressure fluctuates or the pressure setting is
inappropriate.
Adjust the pilot pressure between 400kPa and 700 kPa.
Avoid using air tools branching off the pressure source of the LS-R700 to supply
an stable air.
Setting up a dedicated pressure source for the LS-R700 is recommended.
Malfunction of the pressure switch monitoring the
pilot pressure for the Balance (BAL) valve.
Contact Cosmo for repair.
As a provisional measure, the pressure switch monitoring can be disabled.
Go to: Settings > Common Settings > Special > PSW Monitoring