LS-R700 computes leakage based on the measurement of the pressure difference between the non-leaking
master and the tested part.
K(Ve) is "leak coefficient" which is used for converting measured differential pressure into a flow rate.
For Master --- Master Chamber, Non-leaking part, Tubing or Other.
Leak Master for NG limit check: Flow should be 1.2 ~ 1.5 times
larger than the Leak Limits
Criterion for “Not Leaking” --- Readout = 5Pa ~ 20Pa with relative consistency.
Determine the following conditions before the initial adjustment.
Prepare the following items:
Non-leaking parts (Must be at ambient temperature and dry): 3 or more pieces
Defective (Leaking) parts (Large leak): 1 or more
Calibrator --- For K(Ve) measurement --- Leak Master:
Δ
Pressurization (CHG) = 40s
Equalization (BAL1) = 30s
Stabilization (BAL2) = 5s
Pressurization (CHG) = 60s
Equalization (BAL1) = 0.5s
Stabilization (BAL2) = 30s
Test Pressure UL/LL --- ±10KPa
Perform leak tests 3 times with the known good parts.
Air leak test system is more than likely to be leaking
Pressurization (CHG) = Automatically setup
Equalization (BAL1) = Automatically setup
Stabilization (BAL2) = Automatically setup
Detection (DET) = Fixed to 5s
Use the following settings or perform Automatic Setup for the initial settings.
1. Check the test parameter
2. Spot the leak location by applying soap solution.
2. Check whether the tested part is leaking or clamp force is sufficient.
1. Check the tube and fittings up to the tested part.