Specifications
Operator's Manual A-3
A.2.5 Alarm Conditions
Note:
During multiple-alarm conditions, alarms are presented as follows:
• If two medium-priority alarms occur simultaneously, the alarm status message toggles between the
two.
• If a medium-priority and low-priority alarm occur simultaneously, only the medium-priority alarm is
displayed and only the medium-priority tone is sounded.
• A medium-priority physiological alarm overrides a low-priority alarm.
TableA-5.Alarm Conditions
Priority Rate Color Type Messages
Medium Sounds every 8 seconds
while the alarm condi-
tion persists
Yellow
Steady message, slow
flashing label
Physiological High rSO
2
Low rSO
2
Technical Sensor not connected
Excessive light
Poor signal quality
Preamp not connected
Replace sensor
Interference detected
Sensor unusable
Low Sounds every 16 seconds
while the alarm condi-
tion persists
Steady yellow Technical Battery low
COM port unavailable
Printer unavailable
USB storage full
USB storage error
Auto baseline set