Using the Digital Output Port
Operator's Manual 7-13
3. Set the inputs for the COM port (see Table7-2.).
For additional information about using HyperTerminal™*, select the Help menu in HyperTermi-
nal™*.
Capturing Data Output Using HyperTerminal™*
Data can be captured and stored on the PC using HyperTerminal™*. When the INVOS™ system is
operating in Run function, data can be output in real-time or case history formats. When the
INVOS™ system is operating in Case archive function, the data is output in case history format. To
capture and store the data:
1. On the PC, open the HyperTerminal™* program via the INVOS™ system terminal icon.
2. Select TRANSFER and CAPTURE TEXT. In the CAPTURE TEXT window, enter the file name and directory
to save the file.
3. Select START.
4. On the INVOS™ system, select operating mode and data output format (see below).
5. On the PC, when monitoring is complete, select CALL, DISCONNECT, and close HyperTerminal™*.
In Run Function—Real-Time Data Output
Once the INVOS™ system components and the PC are connected, the INVOS™ Monitor is on and
rSO
2
data is trending, press the HOME key to display the Main screen and perform the fol-
lowing navigation bar options to begin interfacing:
Table7-2.Input COM Port Settings Using HyperTerminal™*
Input Setting
Bits per second 9600 baud
Data bits 8
Parity None
Stop bits 1
Flow control None
BASELINE MENU EVENT MARK ALARM AUDIO ON/OFF NEXT MENU
OUTPUT SELECT USER CONFIGURATION TIME SCALE NEXT MENU
USB DIGITAL OUTPUT PREVIOUS MENU
PC LINK VUELINK PREVIOUS MENU MAIN MENU
CASE HISTORY OUTPUT FORMAT PREVIOUS MENU MAIN MENU