EasyManua.ls Logo

Covidien INVOS 5100C - 6.5.3 Alarm Limits

Covidien INVOS 5100C
178 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Run Function (Patient Monitoring) in Normal Mode
Operator's Manual 6-13
White events reflect relatively routine occurrences in the environment. These are listed in the
order they typically occur. Yellow events mark events clinicians may consider “cautionary”
because they could cause ischemia. These are listed alphabetically. Green events mark interven-
tions taken and are listed alphabetically.
6.5.3 Alarm Limits
WARNING:
Do not set alarm limits to extreme values that can render the INVOS™ system ineffective. Ensure
alarm limits are appropriate for each patient.
Alarm limits can be set and adjusted in the User configuration menu.
Note:
Alarm limits are stored in the INVOS™ system nonvolatile memory. They will reappear as the operator set
them when the INVOS™ system is rebooted.
Note:
The upper and lower alarm limits for each channel are not allowed to cross. When being increased, the
lower alarm limit is allowed to bump the upper alarm limit higher until it reaches 5 units below the upper
Table6-4.Event Codes and Descriptions for Neonatal Intensive Care Unit

Table of Contents

Other manuals for Covidien INVOS 5100C

Related product manuals