If using two different sized syringes in the C and R channels, add the values listed for
th
e smaller syringe size and the larger syringe. If using dilution, add the value from the
column Dilution syringe. If using same size syringes for both channels, skip to the
second table for calculation recommendations.
Syringe size
(m
L)
Values used to calculate Start waste for different sized
syringes (mL)
Smaller syringe Larger syringe Dilution syringe
1 0.05 N/A 0.20
3 0.10 0.20 0.25
5 0.15 0.30 0.35
10 0.30 0.50 0.50
20 0.40 1.40 0.65
30 0.50 2.35 0.65
If the syringes for the C and R channels are the same size, use the start waste value
li
sted in the table below. If using dilution, add the value from the column Dilution
syringe. If using same size syringes for both channels, skip to the second table for
calculation recommendations.
Syringe size (mL)
Values used to calculate St
art waste for same sized
syringes (mL)
Same size syringes Dilution syringe
1 0.20 0.20
3 0.30 0.25
5 0.50 0.35
10 0.80 0.50
20 1.60 0.65
30 2.70 0.65
5 Preparation
5.
1 Parameters
5.1.3 Waste parameters
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 64