Object name
Command Parameter
range
Factory
seng
Data type Units
Comments
Close vent
valve
!C875 1 - decimal - Closes the vent valve for
delayed start and
overrides the current
vent opon. There is no
open vent valve
command but the stop
command (!C852 0) will
clear the override.
Valve 1 type !S877 0 0 decimal - Normally Open Valve
?S877 1 Normally Closed Valve
(power cycle is required)
Valve 2 type !S878 0 0 decimal - Normally Open Valve
?S878 1 Normally Closed Valve
(power cycle is required)
Service status ?V881 32 bits - hex ags Service status word
Controller run
me
?V882 0..999999 - decimal hours Hours run by controller
0..999999 decimal hours Hours unl controller
service due
Pump run
me
?V883 0..999999 - decimal hours Hours run by pump
0..999999 decimal hours Hours unl pump service
due
Pump cycles ?V884 0..65535 - decimal cycles Cycles run by pump
0..65535 decimal cycles Cycles unl pump service
due
Bearing run
me
?V885 0..999999 - decimal hours Hours run by bearing
0..999999 decimal hours Hours unl bearing
service due
Oil cartridge
run me
?V886 0..999999 - decimal hours Hours run by oil cartridge
0..999999 - - -
0..999999 - decimal hours Hours unl oil cartridge
service due
Page 50
B8G000880_D - Installaon