Configuredeviceoptionsandpreferences
OptionDescription
EnabledOperatorscanstartandstoptotalizersand
inventoriesfromthedisplay,ifatleastone
totalizerisconguredasadisplayvariable.
Disabled(default)Operatorscannotstartandstoptotalizersand
inventoriesfromthedisplay.
2.Ensurethatatleastonetotalizerhasbeenconguredasadisplayvariable.
Thisfunctionisaccessedfromatotalizervalueonthedisplay.T oensurethattheoperatorcan
startandstoptotalizersandinventories,atleastonetotalizermustbeshownonthedisplay.
6.2.2EnableordisableTotalizerResetfromthedisplay
Display
OFF-LINEMAINT→OFF-LINECONFG→DISPLAY→TOTALSRESET
ProLinkII
ProLink→Conguration→Display→DisplayOptions→DisplayTotalizerReset
FieldCommunicator
Notavailable
Youcancontrolwhetherornottheoperatorwillbeabletoresettotalizersfromthedisplay.
Restrictions
•TotalizerResetdoesnotapplytoinventories.Operatorscannotresetinventoriesfromthedisplay,
evenifTotalizerResetisenabled.
•Youcannotusethedisplaytoresetalltotalizersasagroup.Fromthedisplay,youmustreset
totalizersindividually.
•
Ifthepetroleummeasurementapplicationisinstalledonyourcomputer,theoperatormustenter
theoff-linepasswordtoperformthisfunction,eveniftheoff-linepasswordisnotenabled.
Procedure
1.EnableordisableTotalizerResetasdesired.
OptionDescription
EnabledOperatorscanresetatotalizerfromthedisplay,
iftheappropriatetotalizerisconguredasa
displayvariable.
Disabled(default)Operatorscannotresettotalizersfromthe
display.
2.Ensurethatthetotalizerstoberesetfromthedisplayhavebeenconguredasdisplayvariables.
104MicroMotion9739MVDTransmitters