Configureprocessmeasurement
CalibrationPressureisthepressureatwhichyoursensorwascalibrated,andthereforedenesthe
pressureatwhichtherewillbenopressureeffect.CalibrationPressureforyoursensorisprovidedon
thesensorcalibrationsheet.Ifthedataisunavailable,enter20PSI.
5.Decidehowpressuredatawillbeprovidedtothetransmitter,andperformtherequiredsetup.
Ifyouwillpollanexternalpressuredevice,setuppollingforpressure.
Ifyouwilluseastaticpressurevalue,setPressureUnitstotheunitsyouareusing,enter
ExternalPressure,andensurethatpollingforpressureisdisabled.
Ifyouwillusedigitalcommunicationsoradirectanalogcurrenttowritepressuredatatothe
transmitter,setPressureUnitstotheunitstobeusedandensurethatpollingforpressure
isdisabled.Thenensurethattheappropriatevalueiswrittentotransmittermemoryat
appropriateintervals.
Ifyouwilluseanexternalmeasurementdevice,congurethemAinputtoExternalPressure.
YoumustalsoenableExternalPressureCompensationandsetthePressureUnitstotheunitssetat
theexternalmeasurementdevice.
5.9Configurethepetroleummeasurementapplication
DisplayNotavailable
ProLinkII
ProLink→Conguration→APISetup
FieldCommunicatorCongure→ManualSetup→Measurements→PetroleumMeasurement
Thepetroleummeasurementparameterscontrolthevaluesthatwillbeusedinthetransmitters
petroleummeasurementapplication.
Thepetroleummeasurementparametersinclude:
APITableType
ThermalExpansionCoefcient(TEC)(ifrequiredbyAPITableType)
ReferenceTemperature(ifrequiredbyAPITableType)
Restriction
Thepetroleummeasurementparametersareavailableonlyifthepetroleummeasurementapplicationhas
beenpurchasedandisenabledonyourtransmitter.
Procedure
1.SelectAPITableType.
2.IfyousetAPITableTypeto53A,53B,53D,or54C,setReferenceTemperaturetotheappropriatevaluefor
yourapplication.Enterthevaluein°C.
3.IfyousetAPITableTypeto6C,24C,or54C,setThermalExpansionCoefcienttotheappropriatevalue
foryourapplication.
4.(Optional)Setthetemperatureunitconguredonthetransmittertothetemperatureunitusedby
yourAPIreferencetable.
ConfigurationandUseManual