AppendixA
Defaultvaluesandranges
Topicscoveredinthisappendix:
♦Defaultvaluesandranges
A.1Defaultvaluesandranges
SeeT ableA-1forthedefaultvaluesandrangesforthemostfrequentlyusedtransmittersettings.
TableA-1Transmitterdefaultvaluesandranges
TypeSettingDefaultRangeComments
FlowdirectionForward
Flowdamping0.8sec0.0–40.96
sec
User-enteredvalueiscorrected
tonearestlowervalueinlist
ofpresetvalues.Forgas
applications,MicroMotion
recommendsaminimumvalue
of2.56.
Massowunitsg/s
Massowcutoff0.0g/sRecommendedsettingis5%
ofthesensor’sratedmaximum
owrate.
Volumeowtype
Liquid
VolumeowunitsL/s
Flow
Volumeowcutoff0/0L/s0.0–xL/s
xisobtainedbymultiplyingthe
owcalibrationfactorby0.2,
usingunitsofL/s.
Massfactor
1
Densityfactor
1
Meterfactors
Volumefactor
1
ConfigurationandUseManual
203