Configureprocessmeasurement
Table5-11APIreferencetables,associatedprocessfluids,andassociatedcalculationvalues
continued
TablenameProcessfluidCTLsourcedata
Reference
temperatureDensityunit
23A
Generalizedcrude
andJP4
Observeddensity
andobserved
temperature
60°F(non-
congurable)
Relativedensity
Range:0.6110to
1.0760
23B
Generalized
products
Observeddensity
andobserved
temperature
60°F(non-
congurable)
Relativedensity
Range:0.6535to
1.0760
23DLubricatingoils
Observeddensity
andobserved
temperature
60°F(non-
congurable)
Relativedensity
Range:8520to
1.1640
24C
Liquidswitha
constantdensity
baseorknown
thermalexpansion
coefcient
User-supplied
referencedensity(or
thermalexpansion
coefcient)
andobserved
temperature
60°F(non-
congurable)
Relativedensity
53A
Generalizedcrude
andJP4
Observeddensity
andobserved
temperature
15°C(congurable)
Basedensity
Range:610to1075
kg/m
3
53B
Generalized
products
Observeddensity
andobserved
temperature
15°C(congurable)
Basedensity
Range:653to1075
kg/m
3
53DLubricatingoils
Observeddensity
andobserved
temperature
15°C(congurable)
Basedensity
Range:825to1164
kg/m
3
54C
Liquidswitha
constantdensity
baseorknown
thermalexpansion
coefcient
User-supplied
referencedensity(or
thermalexpansion
coefcient)
andobserved
temperature
15°C(congurable)
Basedensityinkg/m
3
5.10Configuretheconcentrationmeasurementapplication
DisplayNotavailable
ProLinkII
ProLink→Conguration→CMSetup
FieldCommunicatorCongure→ManualSetup→Measurements→ConcentrationMeasurement
Theconcentrationmeasurementparameterscontrolhowthetransmittercalculatesconcentration
fromtemperatureanddensitydata.
ConfigurationandUseManual
93