Configureprocessmeasurement
tExample:CutoffinteractionwithAOCutoffhigherthanGasStandardVolumeFlowCutoff
Conguration:
mAOutputProcessVariablefortheprimarymAoutput:GasStandardVolumeFlowRate
FrequencyOutputProcessVariable:GasStandardVolumeFlowRate
AOCutofffortheprimarymAoutput:15SLPM(standardlitersperminute)
GasStandardVolumeFlowCutoff:10SLPM
Result:
IftheGasStandardVolumeowratedropsbelow15SLPMbutnotbelow10SLPM:
TheprimarymAoutputwillreportzeroow.
Thefrequencyoutputwillreporttheactualowrate,andtheactualowratewillbeused
inallinternalprocessing.
IftheGasStandardVolumeowratedropsbelow10SLPM,bothoutputswillreportzeroow,
and0willbeusedinallinternalprocessing.
5.5ConfigureFlowDirection
DisplayNotavailable
ProLinkII
ProLink→Conguration→Flow→FlowDirection
FieldCommunicatorCongure→ManualSetup→Measurements→Flow→FlowDirection
FlowDirectioncontrolshowconditionsofforwardowandreverseowaffectowmeasurementand
reporting.
FlowDirectionisdenedwithrespecttotheowarrowonthesensor:
Forwardow(positiveow)movesinthedirectionoftheowarrowonthesensor.
Reverseow(negativeow)movesinthedirectionoppositetotheowarrowonthesensor.
Procedure
SetFlowDirectionasdesired.
5.5.1OptionsforFlowDirection
FlowDirectioncontrolshowtheoutputsreportowandhowthetotalizersandinventoriesincrementtotals.
ConfigurationandUseManual
79