Troubleshooting
10.6Milliampoutputproblems
Table10-6Milliampoutputproblemsandrecommendedactions
ProblemPossiblecausesRecommendedactions
NomAoutput
Wiringproblem
Circuitfailure
Checkthepowersupplyandpower
supplywiring.SeeSection10.9.
ChecktheFaultActionsettings.See
Section10.18.
MeasureDCvoltageacrossoutput
terminalstoverifythattheoutputis
active.
CheckthemAoutputwiring.
ContactMicroMotion.
Looptestfailed
Powersupplyproblem
Incorrectinternal/externalpower
conguration
Outputnotpowered
Wiringproblem
Circuitfailure
Checkthepowersupplyandpower
supplywiring.SeeSection10.9.
ChecktheFaultActionsettings.See
Section10.18.
CheckthemAoutputwiring.
ContactMicroMotion.
mAoutputbelow4
mA
ProcessconditionbelowLRV
LRVandURVarenotsetcorrectly
Faultconditioniffaultindicatorissetto
internalzeroordownscale
Openinwiring
BadmAreceivingdevice
Badoutputcircuit
Verifyprocessconditions.
CheckthesettingsofUpperRangeValue
andLowerRangeValue.SeeSection
10.17.
ChecktheFaultActionsettings.See
Section10.18.
Verifythatthereceivingdevice,and
thewiringbetweenthetransmitterand
thereceivingdevice
ConstantmAoutput
Non-zeroHARTaddress(mAoutput1)
Outputisxedintextmode
Zerocalibrationfailure
ChecktheHARTaddressandLoop
CurrentMode.SeeSection10.14,
Checkthelooptestmode.
CheckHARTburstmodeconguration.
SeeSection10.15.
Ifrelatedtoazerocalibrationfailure,
cyclepowertothemeterandretrythe
zeroingprocedure.
mAoutput
consistentlyoutof
range
Faultconditioniffaultindicatorissetto
upscaleordownscale
LRVandURVarenotsetcorrectly
ChecktheFaultActionsettings.See
Section10.18.
CheckthesettingsofUpperRangeValue
andLowerRangeValue.SeeSection
10.17.
188
MicroMotion9739MVDTransmitters