Configuredeviceoptionsandpreferences
Alarmdataintransmittermemory
Foreachalarmoccurrencethatisposted,informationismaintainedinthreedifferentwaysintransmitter
memory:
•AlertList
•AlertStatistics
•RecentAlerts
Table6-2describesthesethreetypesofalarmdatastructures.
Table6-2Alarmdataintransmittermemory
Transmitteractionifconditionoccurs
AlarmdatastructureContentsClearing
AlertList
Listof:
•Allcurrentlyactivealarms
•Allpreviouslyactivealarmsthat
havenotbeenacknowledged
asdeterminedbythealarmstatus
bits
Clearedandregeneratedwithevery
transmitterpowercycle
AlertStatisticsOnerecordforeachalarm(byalarm
number)thathasoccurredsince
thelastmasterreset.Eachrecord
contains:
•Acountofthenumberof
occurrences
•Timestampsforthemost
recentpostingandclearing
Notcleared;maintainedacross
transmitterpowercycles
RecentAlerts50mostrecentalarmpostingsor
alarmclearings
Notcleared;maintainedacross
transmitterpowercycles
6.6Configureinformationalparameters
DisplayNotavailable
ProLinkII
ProLink→Conguration→Sensor
FieldCommunicatorCongure→ManualSetup→InfoParameters
Theinformationalparameterscanbeusedtoidentifyordescribeyourowmeter.Theyarenotused
intransmitterprocessingandarenotrequired.