Configuredeviceoptionsandpreferences
Table6-1StatusalarmsandStatusAlarmSeverity
continued
Alarm
code
StatusmessageDefault
severityNotesConfigurable?
A105
Slugow
Informational
Yes
A106Burstmodeenabled
InformationalCanbesettoeitherInformationalor
Ignore,butcannotbesettoFault.
Yes
A107Powerresetoccurred
Informational
Normaltransmitterbehavior;
occursaftereverypowercycle.
Yes
A108Event1triggered
Informational
Appliesonlytobasicevents.Yes
A109Event2triggered
Informational
Appliesonlytobasicevents.Yes
A110Frequencyoutputsaturated
InformationalCanbesettoeitherInformationalor
Ignore,butcannotbesettoFault.
Yes
A111
Frequencyoutputxed
InformationalCanbesettoeitherInformationalor
Ignore,butcannotbesettoFault.
Yes
A113
SecondarymAoutputsaturated
InformationalCanbesettoeitherInformationalor
Ignore,butcannotbesettoFault.
Yes
A114
SecondarymAoutputxed
InformationalCanbesettoeitherInformationalor
Ignore,butcannotbesettoFault.
Yes
A115Externalinputerror
Informational
Yes
A116APItemperatureoutsidestandard
range
Informational
Appliesonlytotransmitterswith
thepetroleummeasurement
application.
Yes
A117
APIdensityoutoflimits
Informational
Appliesonlytotransmitterswith
thepetroleummeasurement
application.
Yes
A118
Discreteoutput1xed
InformationalCanbesettoeitherInformationalor
Ignore,butcannotbesettoFault.
Yes
A120
Concentrationmeasurement:
unabletoxcurvedata
Informational
Appliesonlytotransmitterswith
theconcentrationmeasurement
application.
No
A121
Concentrationmeasurement:
extrapolationalarm
Informational
Appliesonlytotransmitterswith
theconcentrationmeasurement
application.
Yes