Procedure
ToperformaD3calibrationorD3andD4calibration,seeFigure9-4.
Figure9-4D3orD3andD4densitycalibrationusingtheFieldCommunicator
9.5Performtemperaturecalibration
Temperaturecalibrationestablishestherelationshipbetweenthetemperatureofthecalibrationuids
andthesignalproducedbythesensor.
Prerequisites
Thetemperaturecalibrationisatwo-partprocedure:temperatureoffsetcalibrationandtemperature
slopecalibration.Thetwopartsmustbeperformedwithoutinterruption,intheordershown.Makesure
thatyouarepreparedtocompletetheprocesswithoutinterruption.
YoumustuseProLinkIItocalibratefortemperature.