Chapter5
Configureprocessmeasurement
Topicscoveredinthischapter:
Characterizetheowmeter
Conguremassowmeasurement
Congurevolumeowmeasurementforliquidapplications
Conguregasstandardvolumeowmeasurement
CongureFlowDirection
Conguredensitymeasurement
Conguretemperaturemeasurement
Congurepressurecompensation
Congurethepetroleummeasurementapplication
Conguretheconcentrationmeasurementapplication
5.1Characterizetheflowmeter
DisplayNotavailable
ProLinkII
ProLink→Conguration→DensityProLink→Conguration→Flow
FieldCommunicatorCongure→ManualSetup→Characterize
Characterizingtheowmeteradjuststhetransmitter’smeasurementalgorithmstomatchtheunique
traitsofthesensoritispairedwith.Thecharacterizationparameters(alsocalledcalibrationparameters)
describethesensorssensitivitytoow,density,andtemperature.Dependingonyoursensortype,
differentparametersarerequired.Valuesforyoursensorareprovidedby
MicroMotiononthesensor
tagorthecalibrationcerticate.
Note
Ifyoursensorandtransmitterwereorderedtogether,thetransmitterhasalreadybeencharacterizedat
thefactory.However,youshouldstillverifycharacterizationparameters.