Configureprocessmeasurement
Procedure
SetDensityCutofftothedesiredvalue.
ThedefaultvalueforDensityCutoffis0.2g/cm
3
.Therangeis0.0g/cm
3
to0.5g/cm
3
.
EffectofDensityCutoffonvolumemeasurement
DensityCutoffaffectsliquidvolumemeasurement.IfthedensityvaluegoesbelowDensityCutoff,thevolume
owrategoesto0.DensityCutoffdoesnotaffectgasstandardvolumemeasurement.Gasstandard
volumevaluesarealwayscalculatedfromthemeasureddensityvalue.
5.7Configuretemperaturemeasurement
Display
OFF-LINEMAINT→OFF-LINECONFG→UNITS→TEMP
ProLinkII
ProLink→Conguration→Density→LowDensityCutoff
FieldCommunicatorCongure→ManualSetup→Measurements→Temperature
Thetemperaturemeasurementparameterscontrolhowtemperaturedatafromthesensorisreported.
Temperaturedataisusedtocompensateowmeasurementfortheeffectoftemperatureonthesensor
tubes.
Thetemperaturemeasurementparametersinclude:
TemperatureMeasurementUnit
TemperatureDamping
5.7.1ConfigureTemperatureMeasurementUnit
Display
OFF-LINEMAINT→OFF-LINECONFG→UNITS→TEMP
ProLinkII
ProLink→Conguration→Temperature→TempUnits
FieldCommunicatorCongure→ManualSetup→Measurements→Temperature→TemperatureUnit
TemperatureMeasurementUnitspeciestheunitthatwillbeusedfortemperaturemeasurement.
Procedure
SetTemperatureMeasurementUnittothedesiredoption.
ThedefaultsettingisDegreesCelsius.
Tip
IfyouareconguringthemAinputtoreceivetemperaturedatafromanexternalmeasurementdevice,
youmustsetthemeasurementunittomatchthetemperaturemeasurementunitattheexternal
measurementdevice.
OptionsforTemperatureMeasurementUnit
ThetransmitterprovidesastandardsetofunitsforTemperatureMeasurementUnit.Differentcommunications
toolsusedifferentlabels.
88
MicroMotion9739MVDTransmitters