Quickstartwiththedisplay
LowerRangeValuespeciesthevalueofMassFlowRatetoberepresentedbyanoutputlevelof
0mAor4mA.
a.ActivateScrolluntilAO14mAorAO10mAappearsonthedisplay,thenactivateSelect.
b.DeneeachcharacterinLowerRangeValue,includingthesign.
UseSelecttohighlightaspeciccharacter.UseScrolltosetthevalueofthecharacter.
c.Whenyouhavesetallcharactersasdesired,activateScrollandSelectsimultaneouslyuntil
SAVE/YES?ashesonthedisplay,thenactivateSelecttowritethevaluetotransmittermemory.
5.SetUpperRangeValue(URV).
UpperRangeValuespeciesthevalueofMassFlowRatetoberepresentedbyanoutputlevelof20mA.
a.ActivateScrolluntilAO120mAappearsonthedisplay,thenactivateSelect.
b.DeneeachcharacterinUpperRangeValue,includingthesign.
UseSelecttohighlightaspeciccharacter.UseScrolltosetthevalueofthecharacter.
c.Whenyouhavesetallcharactersasdesired,activateScrollandSelectsimultaneouslyuntil
SAVE/YES?ashesonthedisplay,thenactivateSelecttowritethevaluetotransmittermemory.
6.ActivateScrolluntilAOEXITappearsonthedisplay,thenactivateSelect.
7.Returnthedisplaytonormaloperation(displayingprocessdata).
a.ActivateScrolluntilIOEXITappearsonthedisplay,thenactivateSelect.
b.ActivateScrolluntilCONFIGEXITappearsonthedisplay,thenactivateSelect.
c.ActivateScrolluntilOFF-LINEEXITappearsonthedisplay,thenactivateSelect.
d.ActivateScrolluntilEXITappearsonthedisplay,thenactivateSelect.
Tip
Anotherwaytoexittheoff-linemenuistorelyonthetime-outfeature.Ifyouhavenotused
theoff-linemenuforapproximately60seconds,thedisplaywillautomaticallyreturntonormal
operation.
2.4Performalooptest
Alooptestisawaytoverifythatthetransmitterandthereceivingdevicearecommunicatingproperly.
AlooptestalsohelpsyouknowwhetheryouneedtotrimmAoutputs.Performingalooptestisnot
arequiredprocedure.However,
MicroMotionrecommendsperformingalooptestforeveryinputor
outputavailableonyourtransmitter.
Procedure
1.TestthemAoutput.
a.ChooseOFFLINEMAINT→SIM→AO1SIM,andselectSET4MAoranothermAoutputvalue.
Dotstraversethedisplaywhiletheoutputisxed.
b.ReadthemAcurrentatthereceivingdeviceandcompareittothetransmitteroutput.
Thereadingsdonotneedtomatchexactly.Ifthevaluesareslightlydifferent,youcancorrect
thediscrepancybytrimmingtheoutput.