Measurementsupport
IfyouperformtheD3andD4calibration,notethefollowing:
DonotperformtheD1andD2calibration.
PerformtheD3calibrationifyouhaveonecalibrateduid.
PerformboththeD3andD4calibrationsifyouhavetwocalibrateduids(otherthanairand
water).Thecalibrationsmustbeperformedwithoutinterruption,intheordershown.Makesure
thatyouarepreparedtocompletetheprocesswithoutinterruption.
9.4.1PerformaD3andD4densitycalibrationusingProLinkII
Prerequisites
Duringdensitycalibration,thesensormustbecompletelylledwiththecalibrationuid,and
owthroughthesensormustbeatthelowestrateallowedbyyourapplication.Thisisusually
accomplishedbyclosingtheshutoffvalvedownstreamfromthesensor,thenllingthesensor
withtheappropriateuid.
ForD3densitycalibration,theD3uidmustmeetthefollowingrequirements:
Minimumdensityof0.6g/cm
3
Minimumdifferenceof0.1g/cm
3
betweenthedensityoftheD3uidandthedensityofwater.
ThedensityoftheD3uidmaybeeithergreaterorlessthanthedensityofwater.
ForD4densitycalibration,theD4uidmustmeetthefollowingrequirements:
Minimumdensityof0.6g/cm
3
Minimumdifferenceof0.1g/cm
3
betweenthedensityoftheD4uidandthedensityofthe
D3uid.ThedensityoftheD4uidmustbegreaterthanthedensityoftheD3uid.
Minimumdifferenceof0.1g/cm
3
betweenthedensityoftheD4uidandthedensityofwater.
ThedensityoftheD4uidmaybeeithergreaterorlessthanthedensityofwater.
Beforeperformingthecalibration,recordyourcurrentcalibrationparameters.Youcandothisby
savingthecurrentcongurationtoaleonthePC.Ifthecalibrationfails,restoretheknownvalues.
Procedure
ToperformaD3calibrationorD3andD4calibrationusingProLinkII,seeFigure9-3.
172MicroMotion9739MVDTransmitters