Configureprocessmeasurement
5.3.3ConfigureVolumeFlowCutoff
DisplayNotavailable
ProLinkII
ProLink→Conguration→Flow→VolFlowCutoff
FieldCommunicatorCongure→ManualSetup→Measurements→Flow→VolumeFlowCutoff
VolumeFlowCutoffspeciesthelowestvolumeowratethatwillreportedasmeasured.Allvolume
owratesbelowthiscutoffwillbereportedas0.
Procedure
SetVolumeFlowCutofftothedesiredvalue.
ThedefaultvalueforVolumeFlowCutoffis0.0L/s.Thelowerlimitis0.Theupperlimitisthesensor’sow
calibrationfactor,inunitsofL/s,multipliedby0.2.
InteractionbetweenVolumeFlowCutoffandAOCutoff
VolumeFlowCutoffdenesthelowestliquidvolumeowvaluethatthetransmitterwillreportasmeasured.
AOCutoffdenesthelowestowratethatwillbereportedviathemAoutput.IfmAOutputProcessVariableis
settoVolumeFlowRate,thevolumeowratereportedviathemAoutputiscontrolledbythehigherof
thetwocutoffvalues.
VolumeFlowCutoffaffectsbothvolumeowvaluesreportedviaoutputsandvolumeowvaluesusedin
othertransmitterbehavior(e.g.,eventsdenedonvolumeow).
AOCutoffaffectsonlyowvaluesreportedviathemAoutput.
tExample:CutoffinteractionwithAOCutofflowerthanVolumeFlowCutoff
Conguration:
mAOutputProcessVariablefortheprimarymAoutput:VolumeFlowRate
FrequencyOutputProcessVariable:VolumeFlowRate
AOCutofffortheprimarymAoutput:10liters/second
VolumeFlowCutoff:15liters/second
Result:Ifthemassowratedropsbelow15liters/second,volumeowwillbereportedas0,and0
willbeusedinallinternalprocessing.
tExample:CutoffinteractionwithAOCutoffhigherthanVolumeFlowCutoff
Conguration:
mAOutputProcessVariablefortheprimarymAoutput:VolumeFlowRate
FrequencyOutputProcessVariable:VolumeFlowRate
AOCutofffortheprimarymAoutput:15liters/second
VolumeFlowCutoff:10liters/second
ConfigurationandUseManual
73

Table of Contents