5.6.2Configureslugflowparameters
DisplayNotavailable
ProLinkII
ProLink→Conguration→Density→SlugHighLimit
ProLink→Conguration→Density→SlugLowLimit
ProLink→Conguration→Density→SlugDuration
FieldCommunicatorCongure→ManualSetup→Measurements→Density→SlugLowLimit
Congure→ManualSetup→Measurements→Density→SlugHighLimit
Congure→ManualSetup→Measurements→Density→SlugDuration
Theslugowparameterscontrolhowthetransmitterdetectsandreportstwo-phaseow.
Procedure
1.SetSlugLowLimittothelowestdensityvaluethatisconsiderednormalinyourprocess.
Valuesbelowthiswillcausethetransmittertoperformtheconguredslugowaction.Typically,
thisvalueisthelowestdensityvalueinthenormalrangeofyourprocess.
Tip
Gasentrainmentcancauseyourprocessdensitytodroptemporarily.T oreducetheoccurrence
ofslugowalarmsthatarenotsignicanttoyourprocess,setSlugLowLimitslightlybelowyour
expectedlowestprocessdensity.
YoumustenterSlugLowLimiting/cm
3
,evenifanotherunithasbeenconguredfordensity
measurement.
ThedefaultvalueforSlugLowLimitis0.0g/cm
3
.Therangeis0.0g/cm
3
to10.0g/cm
3
.
ConfigurationandUseManual