Cerabar M PMC51, PMP51, PMP55
14 Endress+Hauser
PMP51 and PMP55 – metallic process isolating diaphragm for gauge pressure
Sensor Maximum sensor measuring
range
Lowest
calibratable
span (preset at
factory)
1)
MWP OPL Vacuum resistance
2)
Option
3)
lower (LRL) upper (URL) Silicone oil/
Inert oil/
Synthetic oil
[bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar
abs
(psi
abs
)]
400 mbar (6 psi) –0.4 (–6) +0.4 (+6) 0.02 (0.3) 4 (60) 6 (90)
0.01/0.04/0.01
(0.15/0.6/0.15)
1F
1 bar (15 psi) –1 (–15) +1 (+15) 0.05 (1) 6.7 (100) 10 (150) 1H
2 bar (30 psi) –1 (–15) +2 (+30) 0.1 (1.5) 13.3 (200) 20 (300) 1K
4 bar (60 psi) –1 (–15) +4 (+60) 0.2 (3) 18.7 (280.5) 28 (420) 1M
10 bar (150 psi) –1 (–15) +10 (+150) 0.5 (7.5) 26.7 (400.5) 40 (600) 1P
40 bar (600 psi) –1 (–15) +40 (+600) 2 (30) 100 (1500) 160 (2400) 1S
100 bar
(1 500 psi)
–1 (–15) +100 (+1500) 5 (75) 100 (1500) 400 (6000) 1U
400 bar
(6 000 psi)
–1 (–15) +400 (+6000) 20 (300) 400 (6000) 600 (9000) 1W
1) Highest factory-configurable turn down: 20:1, higher available on request or can be set in device.
2) The vacuum resistance applies to the measuring cell at reference conditions. A ceramic process isolating diaphragm is recommended for
applications in the limit range. The pressure and temperature application limits of the selected filling oil must also be observed for the PMP55
→  115.
3) Product Configurator, order code for "Sensor range"
PMP51 and PMP55 – metallic process isolating diaphragm for absolute pressure
Sensor Maximum sensor
measuring
1)
Lowest
calibratable
span (preset at
factory)
2)
MWP OPL Vacuum resistance
3)
Option
4)
lower (LRL) upper (URL) Silicone oil/
Inert oil/
Synthetic oil
[bar
abs
(psi
abs
)]
[bar
abs
(psi
abs
)] [bar (psi)] [bar
abs
(psi
abs
)]
[bar
abs
(psi
abs
)]
[bar
abs
(psi
abs
)]
400 mbar (6 psi) 0 +0.4 (+6) 0.02 (0.3) 4 (60) 6 (90)
0.01/0.04/0.01
(0.15/0.6/0.15)
2F
1 bar (15 psi) 0 +1 (+15) 0.05 (1) 6.7 (100) 10 (150) 2H
2 bar (30 psi) 0 +2 (+30) 0.1 (1.5) 13.3 (200) 20 (300) 2K
4 bar (60 psi) 0 +4 (+60) 0.2 (3) 18.7 (280.5) 28 (420) 2M
10 bar (150 psi) 0 +10 (+150) 0.5 (7.5) 26.7 (400.5) 40 (600) 2P
40 bar (600 psi) 0 +40 (+600) 2 (30) 100 (1500) 160 (2400) 2S
100 bar
(1 500 psi)
0 +100 (+1500) 5 (75) 100 (1500) 400 (6000) 2U
400 bar
(6 000 psi)
0 +400 (+6000) 20 (300) 400 (6000) 600 (9000) 2W
1) PMP55: Within the sensor measuring range, the minimum upper range value of 80 mbar
abs
(1.16 psi
abs
) must be observed.
2) Highest factory-configurable turn down: 20:1, higher available on request or can be set in device.
3) The vacuum resistance applies to the measuring cell at reference conditions. A ceramic process isolating diaphragm is recommended for
applications in the limit range. The pressure and temperature application limits of the selected filling oil must also be observed for the PMP55
→  115.
4) Product Configurator, order code for "Sensor range"