10 Block SPECIAL FUNCTION Device Functions Proline Promass 83 MODBUS RS485
126 Endress+Hauser
EXPANSION
COEFFICIENT
SQUARE CARRIER
FLUID
MODBUS register:
Data type:
Access:
7003
2212
Float
read/write
! Note!
Function is not available unless % MASS / % VOLUME or % BLACK-LIQUOR was
selected in the DENSITY FUNCTION function (7000).
For entering the fluid-specific expansion coefficient for the carrier fluid for
nonlinear temperature curves. This value is required for temperature-
compensated calculation of the target fluid content in a two-phase fluid.
User input: 5-digit floating-point number, including unit and sign
Factory setting: 0.0000 e-6 [1/K
2
]
REFERENCE
DENSITY TARGET
FLUID
MODBUS register:
Data type:
Access:
7004
2214
Float
read/write
! Note!
Function is not available unless % MASS / % VOLUME or % BLACK-LIQUOR was
selected in the DENSITY FUNCTION function (7000).
Use this function to enter the reference density (density at reference temp.) of the
target fluid. This value is required for temperature-compensated calculation of the
target fluid content in a two-phase fluid.
User input: 5-digit floating-point number, including units
Factory setting: 1.0000 kg/l
! Note!
Carrier fluid = transporting liquid, (e.g. water)
Target fluid = material transported (e.g. lime powder)
The appropriate unit is taken from the function UNIT REFERENCE DENSITY
(0421) (see Page 18).
EXPANSION
COEFFICIENT
LINEAR TARGET
FLUID
MODBUS register:
Data type:
Access:
7005
2216
Float
read/write
! Note!
Function is not available unless % MASS / % VOLUME or % BLACK-LIQUOR was
selected in the DENSITY FUNCTION function (7000).
For entering the fluid-specific expansion coefficient for the target fluid for linear
temperature curves. This value is required for temperature-compensated
calculation of the target fluid content in a two-phase fluid.
User input: 5-digit floating-point number, including unit and sign
Factory setting: 0.5000 e-3 [1/K]
EXPANSION
COEFFICIENT
SQUARE TARGET
FLUID
MODBUS register:
Data type:
Access:
7006
2218
Float
read/write
! Note!
Function is not available unless % MASS / % VOLUME or % BLACK-LIQUOR was
selected in the DENSITY FUNCTION function (7000).
For entering the fluid-specific expansion coefficient for the target fluid for
nonlinear temperature curves. This value is required for temperature-
compensated calculation of the target fluid content in a two-phase fluid.
User input: 5-digit floating-point number, including unit and sign
Factory setting: 0.0000 e-6 [1/K
2
]
LINEAR
EXPANSION
COEFFICIENT
MODBUS register:
Data type:
Access:
7007
2222
Float
read/write
! Note!
Function is not available unless °BAUME < 1SG, °BAUME >1SG, °API, °PLATO,
°BALLING or °BRIX was selected in the DENSITY FUNCTION function (7000).
For entering the fluid-specific expansion coefficient (for linear temperature
curves), to calculate the temperature-compensated density functions.
User input: 5-digit floating-point number, including unit and sign
Factory setting: 0.5000 e-3 [1/K]
Function description
SPECIAL FUNCTION DENSITY FUNCTIONS CONFIGURATION