Device Functions Proline Promass 83 MODBUS RS485 7 Block OUTPUTS
Endress+Hauser 59
7.1 Group CURRENT OUTPUT
7.1.1 Function group CONFIGURATION
OUTPUTS E → CURRENT OUTPUT 1 EAA → CONFIGURATION 400
Function description
OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION
ASSIGN CURRENT
OUTPUT 1
MODBUS register:
Data type:
Access:
4000
5801
Integer
read/write
For assigning a measured variable to the current output.
Options:
0 = OFF
2 = MASS FLOW
5 = VOLUME FLOW
6 = CORRECTED VOLUME FLOW
7 = DENSITY
8 = REFERENCE DENSITY
9 = TEMPERATURE
Advanced selection: with the optional SW package CONCENTRATION
12 = TARGET MASS FLOW
13 = % TARGET MASS FLOW
14 = TARGET VOLUME FLOW
15 = % TARGET VOLUME FLOW
16 = CORRECTED TARGET VOLUME FLOW
17 = CARRIER MASS FLOW
18 = % CARRIER MASS FLOW
19 = CARRIER VOLUME FLOW
20 = % CARRIER VOLUME FLOW
21 = CORRECTED CARRIER VOLUME FLOW
22 = % BLACK LIQUOR
23 = ° BAUME
25 = ° API
26 = ° PLATO
27 = ° BALLING
28 = ° BRIX
29 = OTHER (_ _ _ _ flexible concentration)
Advanced selection: with the optional SW package ADVANCED DIAGNOSIS:
58 = MASS FLOW DEVIATION
59 = DENSITY DEVIATION
60 = REFERENCE DENSITY DEVIATION
61 = TEMPERATURE DEVIATION
62 = TUBE DAMPING DEVIATION
63 = ELECTRODYNAMIC SENSOR DEVIATION
86 = OPERATING FREQUENCY FLUCTUATION DEVIATION
87 = TUBE DAMPING FLUCTUATION DEVIATION
Factory setting: MASS FLOW
! Note!
If you select OFF, the only function shown in the CONFIGURATION function
group is this function, in other words ASSIGN CURRENT OUTPUT 1 (4000).