4 Block QUICK SETUP Device Functions Proline Promass 83 MODBUS RS485
24 Endress+Hauser
4.2 “Batching” Quick Setup menu
!
Note!
This function is only available when the additional “batching” software is installed in the measuring
device (order option). You can order this software from Endress+Hauser as an accessory at a later
date (see Operating Instructions).
This Quick Setup guides the user systematically through all the device functions
that have to be adjusted and configured for batching operation. These basic settings allow simple
(one step) batching processes.
Additional settings, e.g. for multi-stage batching processes, have to be made via the function matrix
itself.
A0004644-en
Esc
E
+
-
XXX.XXX.XX
+
B
6402
6603
6404
7200
7201
7203
7204
1005
4700
4780
7240
+
E
E
ON-Value
Low flow cut off
Flow
damping
Pressure shock
suppression
Batch
Selector
Batch
Name
Batch
Quantity
Fix
Compensation Quantity
Assign
Relay
Terminal No.
Max. Batch
Time
QS
Batching/Dosing
YES NO
Batch
Supervision?
YES NO
Autom. Configuration
Display?
Automatic parameterization
of the display
Quick Setup
Quit Quick Setup
Locked Access Customer
PRESET
Batch quantity
Relay 1 Relay 2
Select
Output?
HOME-POSITION