Device Functions Proline Promass 83 MODBUS RS485 4 Block QUICK SETUP
Endress+Hauser 25
"
Caution!
The “Batching” Quick Setup sets certain device parameters for discontinuous measurement
operation. If the measuring device is used for continuous flow measurement at a later time, we
recommend at you rerun the “Commissioning” and/or “Pulsating Flow” Quick Setup.
!
Note!
• The display returns to the function QUICK SETUP BATCHING/DOSING (1005) if you press the
ESC key combination during parameter interrogation.
• At the start of the Setup, general device parameters are optimally configured for measuring signal
processing and output response.
• Then you can enter the specific batching parameters, starting with the options list “Batching 1
to 6”. In this way, by running through the Setup menu a number of times, up to six different
batching parameter sets (incl. special naming) can be created and called up as necessary.
• In order to enjoy full functionality, it is advisable to let the display parameters be set
automatically. This means that the lowest display line is parameterized as the batching menu.
Softkeys are displayed which can be used to start or stop the batching process in the HOME
position. In this way, the measuring device can be fully deployed as a “batch controller”.
• The “PRESET batch quantity” prompt is used to specify the type of access to the batch quantity:
– “Access customer” → The batch quantity can be modified via local operation (softkeys),
without having to first enter the defined private code.
– “Locked” → The current batch quantity can only be read but cannot be edited without first
entering the defined private code.