11 Block SUPERVISION Device Functions Proline Promass 83 MODBUS RS485
174 Endress+Hauser
11.2.4 Function group F-CHIP
SUPERVISION J SYSTEM JAA
VERSION INFO JCA DEVICE 810
SENSOR 820
AMPLIFIER 822
F-CHIP 824
Function description
SUPERVISION VERSION INFO F-CHIP
STATUS F-CHIP
MODBUS register:
Data type:
Access:
8240
7057
Integer
read
Use this function to check whether an F-CHIP is installed and which software
options are available.
Display:
0 = NO F-CHIP HW
1 = F-CHIP OK
2 = DEMONSTRATION
3 = SERVICE&ANALYSIS
4 = P-TYPE F-CHIP
5 = TYPE UNKNOWN
6 = SER. NUM. WRONG
7 = FAILURE
8 = CRITICAL FAIL.
9 = DEVICE ID
SYSTEM OPTIONS
MODBUS register:
Data type:
Access:
8241
7058
Integer
read
! Note!
Function is not available unless the measuring device is equipped with an
F-CHIP.
The software options available in the measuring device appear on the display.
Display:
0 = NO ENTRY
4 = DENSITY FUNCTION
5 = BATCH FUNCTION
6 = ADV. DIAGNOSTIC
7 = VISCOSITY
SW REV. NO.
F-CHIP
MODBUS register:
Data type:
Access:
8244
7059
String (16)
read
! Note!
The F-CHIP must be available in order to access this function.
Use this function to view the software revision number of the F-CHIP.