5 Block USER INTERFACE Device Functions Proline Promass 83 MODBUS RS485
36 Endress+Hauser
5.1.3 Function group OPERATION
USER INTERFACE C CONTROL CAA BASIC CONFIGURATION 200
UNLOCKING/LOCKING 202
OPERATION 204
Function description
USER INTERFACE CONTROL OPERATION
TEST DISPLAY
MODBUS register:
Data type:
Access:
2040
2513
Integer
read/write
For testing the operability of the local display and its pixels.
Options:
0 = OFF
1 = ON
Factory setting: OFF
Test sequence:
1. Start the test by selecting “ON”.
2. All pixels of the main line, additional line and information line are darkened
for minimum 0.75 seconds.
3. Main line, additional line and information line show an “8” in each field for
minimum 0.75 seconds.
4. Main line, additional line and information line show a “0” in each field for
minimum 0.75 seconds.
5. Main line, additional line and information line show nothing (blank display)
for minimum 0.75 second.
When the test is completed, the local display returns to its initial state and the
setting changes to “OFF”.