Device Functions Proline Promass 83 MODBUS RS485 5 Block USER INTERFACE
Endress+Hauser 47
5.4 Group INFORMATION LINE
5.4.1 Function group CONFIGURATION
USER INTERFACE C → CONTROL CAA
→
MAIN LINE CCA
→
ADDITIONAL LINE CEA
→
INFORMATION LINE CGA → CONFIGURATION 260
Function description
USER INTERFACE → INFORMATION LINE → CONFIGURATION
ASSIGN
MODBUS register:
Data type:
Access:
2600
2537
Integer
read/write
For assigning a value to be displayed to the information line (bottom line in the
local display). This value is displayed during normal operation.
Options: standard
0 = OFF
1 = MASS FLOW
2 = MASS FLOW IN %
3 = VOLUME FLOW
4 = VOLUME FLOW IN %
5 = CORRECTED VOLUME FLOW
6 = CORRECTED VOLUME FLOW IN %
7 = DENSITY
8 = REFERENCE DENSITY
9 = TEMPERATURE
10 = MASS FLOW BARGRAPH IN %
11 = VOLUME FLOW BARGRAPH IN %
12 = CORRECTED VOLUME FLOW BARGRAPH IN %
15 = ACTUAL CURRENT
18 = ACTUAL FREQUENCY
20 = TOTALIZER 1
21 = TOTALIZER 2
22 = TOTALIZER 3
23 = TAG NAME
24 = OPERATING/SYSTEM CONDITIONS
26 = DISPLAY FLOW DIRECTION
Factory setting: OPERATING/SYSTEM CONDITIONS
Advanced selection: with the optional SW package BATCHING
60 = BATCHING KEYS (softkeys on the local display)
! Note!
• If you select the BATCHING OPERATING KEYS, the multiplex display
functionality is not available in the information line.
• For the functionality and display examples of the Batching menu, see Operating
Instructions BA0107D, “Operation” section.
(continued on next page)